Thursday, December 11, 2008

pH and dissolved oxygen probe cleaning

Question:
Which Alconox brand cleaner is well suited for cleaning pH and dissolved oxygen probes?

Answer:
Dissolved oxygen and pH probes used in fermentation vessels can get buildup from the fermentation media that degrades the probe performance and increases the likelihood of batch to batch contamination. Even CIP systems cannot always reliably clean the port crevices and o-rings. This is a particular problem with proteins found in pharmaceutical, biotech, cheese and brewery manufacturing. Use a 1% Tergazyme solution that does not exceed 130 deg F (55 deg C), preferably around 120 deg F (50 deg C). Clean manually with a small brush. Rinse thoroughly with water of suitable purity for your industry. O-Rings and other removable parts can be soaked in a warm (120 deg F, 50 deg C) solution for 20 minutes prior to final brushing and rinsing.

To view the technical bulletin for Tergazyme, Enzyme Active Cleaner, please click here.

No comments:

Post a Comment